RMT-01 Running Metal Wire Thermometer
Overview

For measuring the temperature of running copper and ultra-fine wire
Measures the temperature of running metal wire using a fine-wire Type K thermocouple. Its small heat capacity and low-friction pulley make it suitable even for metal wire of φ0.1 mm or less.
The measuring range is 10–250°C, and the measured temperature is shown on the unit's LCD.
Product highlights
Features
- Measures the temperature of running metal wire with the small heat capacity of a fine-wire Type K thermocouple.
- Power supply to the pulley and transfer of temperature data are non-contact, reducing rotational friction.
- Prevents thermocouple short-circuit failures caused by copper dust and other deposits.
- The compact, lightweight, battery-powered design can be used wherever it is installed.
- Can be calibrated with the dedicated RMR-01 temperature calibrator.
RMT-01 specifications
Product Specification
RMT-01 Running Metal Wire Thermometer
| Model | RMT-01 | |
|---|---|---|
| Sensor | Type K thermocouple | |
| Measuring temperature range | 10–250°C | |
| Display | 4-digit 7-segment LCD | |
| Display update rate | 1/s | |
| Power supply | Four AA batteries and AC adapter | |
| Power consumption | 6 V, 0.4 A (dry cells) / 5 V, 0.5 A (rechargeable batteries) | |
| CAD files | ||
Reference: battery life with dry cells is 4 hours.
RMR-01 Temperature Calibrator
Temperature Calibrator

A dedicated device for calibrating under conditions close to actual operation
A made-to-order device for calibrating the RMT-01. The Type K thermocouple pulley is run in contact with a copper plate heater set to a given temperature to check its characteristics while running.
The displayed temperature is verified with a dedicated thermometer traceable to national standards.
RMR-01 specifications
Calibrator Specification
RMR-01 Temperature Calibrator (Made to Order)
| Model | RMR-01 | |
|---|---|---|
| Compatible models | RMT-01 | |
| Heater material | Copper | |
| Heater plate thickness | 0.5mm | |
| Set temperature range | 0–150°C | |
| Heater temperature accuracy | Set value ±3°C (at a set temperature of 100°C) | |
| Temperature variation within the heater | ±3℃ | |
| Power supply | AC100V | |
| Power consumption | 200 W (heater on) | |
| CAD files | ||
The RMR-01 is made to order.
How to Use the Calibrator
Calibration Guide
Watch the calibration procedure on video
You can view how to use the RMR-01 on our public website.
Documents for Evaluation
- Download
For case studies and products,
please feel free to contact us.