FAQ
Questions About Laser Outer Diameter Gauges
Questions About Laser Outer Diameter Gauges
-
How do I remove the display unit cover?
-
- If the unit is mounted in a panel, first disconnect all panel wiring, remove the mounting brackets, and take the unit out of the panel.
- Before removing the cover, remove any short pieces that short-circuit terminals.
- Remove the nylon spacers from the rear panel, and the cover comes off.

-
How do I check the display unit version?
-
The version number is the value enclosed in "-" that appears after "8.8.8.8.8." is shown at power-on.
-
How far can the cable between the detector and display unit be extended?
-
The standard cable supplied between the detector and display unit is 5 m.
If you need a longer cable, it can be extended up to 30 m as an option.
We stock cables in six lengths: 5 m, 10 m, 15 m, 20 m, 25 m, and 30 m. For other lengths, please contact our Sales Department.
To extend beyond 30 m, use a relay box (LDM-201 / LDM-202).
The LDM-201 / LDM-202 is a compact box powered by the display unit, so no AC power supply is needed.
With this relay box, the cable can be extended to about 120 m.
-
What happens if I combine a detector and display unit with different serial numbers?
-
The detector and display unit of a laser outer diameter gauge are calibrated as a pair. Wherever possible, do not change the combination, and use units with the same serial number.
If you must change the combination, note the following.- Use a detector shipped with an LDM-110 series unit with an LDM-110 series display unit, and a detector shipped with an LDM-100 series unit with an LDM-100 series display unit.
- Detectors shipped with the LDM-110 series have a laser ON monitor LED.
- A detector shipped with the LDM-110 series cannot measure when connected to the LDM-100 series.
- The "E-00" error message appears, or the unit stops with "0000" shown.
- If the detector is the same type as the display unit's original pair, the values shown will be roughly accurate, but the specified accuracy cannot be guaranteed.
- For example, on the LDM-303H with an accuracy of ±2 μm, changing the combination may cause a deviation of up to about 10 μm. In this case, we recommend recalibrating before use.
- A detector shipped with the LDM-100 series can measure when connected to an LDM-110, but using this combination for a long time may damage either the detector or the display unit (LDM-110).
- Even if one of them is damaged, normal measurement is still possible, but the units will be unable to measure when returned to their original combination.
- Combining with a detector of a different type gives erratic values. In this case, calibration is always required.
For how to calibrate the laser outer diameter gauge, see the instruction manual supplied with it.
- Use a detector shipped with an LDM-110 series unit with an LDM-110 series display unit, and a detector shipped with an LDM-100 series unit with an LDM-100 series display unit.
-
What is the relationship between actual measurement sampling and the averaging count?
-
The detector of a laser outer diameter gauge scans the product 400 times per second.
In other words, it samples the measured value 400 times per second, so the sampling interval is 2.5 ms.
The averaging count of a laser outer diameter gauge is a parameter that sets how many of these samples are collected and averaged.
If the averaging count is set to 256, the display shows the simple average: the sum of 256 samples divided by 256. The display therefore updates every 0.64 s (= 256 × 2.5 ms).
However, the recorder output uses a moving average, so it is updated once every 2.5 ms.When a pipe runs at a line speed of 70 m/min, it moves 1167 mm per second (= 70 m / 60 s).
The sampling interval along the length of the pipe is therefore 2.92 mm (= 1167 mm / 400 scans).
If the averaging count is set to 256 for this pipe, the average is calculated every 0.64 s, so each value is the average over a 747 mm length of pipe (= 1167 mm × 0.64 s).
Measurement accuracy (stability) improves as the averaging count increases, and the relationship follows a Gaussian distribution.
Therefore, when accuracy is required, the averaging count must be set reasonably high.
The distribution is Gaussian because of uneven scanning motor rotation, rotating mirror accuracy, digitization errors in the electronic circuits, and vibration of the workpiece.
-
How does workpiece vibration affect measurement accuracy?
-
The scanning speed of the laser beam is detector measuring range ÷ 400 µs, so the time the laser beam actually takes to cross the workpiece is workpiece diameter ÷ scanning speed.
The maximum measurement error caused by workpiece vibration in one scan is the time the laser beam takes to cross the workpiece ÷ the vibration period of the workpiece × the vibration amplitude of the workpiece.
In summary, the effect of vibration on measurement (maximum error) is workpiece diameter ÷ measuring range ÷ vibration period × vibration amplitude × 400 µs.
When measuring a workpiece with a diameter of 60 mm vibrating at a frequency of 10 Hz and an amplitude of 2 mm with an LDM-304H (measuring range: 90 mm), the maximum error caused by vibration is calculated as follows.
60 mm diameter ÷ 90 mm measuring range ÷ 0.1 s vibration period × 2 mm amplitude × 400 µs = 5.33 µm
This is the maximum error caused by vibration.
In actual measurement, the error is +5.33 µm when the beam travels in the same direction as the workpiece moves, and -5.33 µm when they move in opposite directions.
The maximum error above is the largest error that can occur in a single scan; increasing the averaging count averages out the error.
In addition, with a higher averaging count, the vibration direction and scanning direction repeatedly match and oppose each other, so the errors cancel out.
-
Are there any restrictions on the mounting orientation of the detector?
-
There are no restrictions. The detector can be mounted in any orientation.
However, if the glass on either the emitter or receiver side faces upward, dust and water droplets collect on it more easily, so take particular care to keep the glass clean.
-
How accurate is measurement outside the specified area where the specified accuracy is not guaranteed?
-
Outside the specified area, accuracy is about 2 to 3 times the accuracy guaranteed within the specified area.
Accuracy tends to be closer to the guaranteed value near the specified area and to drift further from it the farther you move away; near the limit of the measuring area, it degrades to about 3 times the guaranteed accuracy.
However, this varies from unit to unit. For one LDM-304H, the accuracy is ±5 µm in the specified 60 mm square area centered on the axis, ±7 µm in a 70 mm square area, ±10 µm in an 80 mm square area, and ±15 µm in a 90 mm square area.
-
Can objects in water be measured through an acrylic water tank or similar?
-
Yes.
However, bubbles or impurities in the water interfere with measurement, even if they are too small to see with the naked eye.
-
What does the hold input circuit look like?
-
It is as shown in the diagram.
The signal is received by a photocoupler, but it is not isolated from the internal circuit.
To prevent noise, do not bundle the wiring together with power lines.
Use twisted-pair wire if possible, and keep the wiring as short as possible.
-
How far apart can the emitter and receiver of a separate-type detector be?
-
The measurable emitter-to-receiver distances for separate types are shown in the table below.
The distances in the table are measurable distances and do not guarantee accuracy. Accuracy decreases somewhat as the distance increases.
Accuracy depends on the distance and the application, so please contact us for details.
Model Maximum measurable distance LDM-303H-SP 400mm LDM-304H-SP 1000mm LDM-305H-SP 3500mm LDM-306H-SP 4000mm
-
Can you give specific examples of how to use the signal splitter unit?
-
Ex. 1) Measuring the diameter of a product and its "runout" or detection position at the same time
This example quantifies the runout of a rotating product.
The signal splitter unit is useful when you want to use two display units, one showing the diameter and the other showing the runout or detection position.

Ex. 2) Measuring the diameters of multiple products at the same time with one detector
This example measures the diameters of two products of different sizes (A and B) with one detector.
The signal splitter unit is also useful when you want to display each diameter at the same time.

There are many other application examples. For details, please contact our Sales Department.
-
What is the difference between the LDM-110/110EX and the LDM-130/130EX?
-
See the comparison table below. Shaded cells show the differences between the two.
Product LDM-110 / 110 EX LDM-130 / 130 EX Measurement averaging count 1 to 2048 (12 steps) Display digits Up to 5 digits Center value setting 5-digit digital switch 5-digit 7-segment LED display Upper tolerance setting 4-digit digital switch 4-digit 7-segment LED display Lower tolerance setting 4-digit digital switch 4-digit 7-segment LED display Maximum / minimum value Displays the maximum or minimum measured value for each averaging cycle USB connector Load / save settings to a USB memory stick Center position monitor 8-position display on an LED bar graph Recorder output ±10 V, selectable from 1 µV to 1 mV Upper / lower limit contact output Upper: 1a, lower: 1a, 250 V AC (resistive load) Error contact output 1b (closed when power is off), 250 V AC, 2 A (resistive load) External communication RS-232 interface as standard RS-485 interface, Modbus (RTU) 2-wire, as standard
CC-Link interface available (option) EtherNet/IP available (option) BCD output BCD output [6 digits] (option)
Error contact output 400 scans/s 400 scans/s or 2000 scans/s External dimensions W144×H72×D191mm
Questions About the BCD Option
-
Can the strobe pulse width of the BCD board be changed? (LDM-110/110EX/210)
-
Yes, but the method differs by model. For details, please ask us through our inquiry form.
-
What are the specifications of the BCD output circuit, and how is it wired to the receiving device?
-
The BCD output circuit specifications are as follows.
- Open collector, negative logic
- Maximum output current when the output transistor is ON: 40 mA DC
- Maximum external power supply voltage that can be connected: +30 V DC
If the receiving device has voltage inputs, connect it as shown in the diagram below.
If the receiving BCD device has no built-in pull-up resistors, add them.

If the receiving device has current inputs (photocoupler inputs), the connection shown below is typical.
If the device has a built-in BCD power supply, the external power supply can be omitted.

In either case, choose an external power supply rated to deliver enough current when all bits are ON.
-
How far can the BCD signal wiring be extended?
-
It depends on the noise environment at the installation site, how the wiring is shielded, and how it is routed, so we cannot give a single answer. Based on the experience of our technical staff, up to about 50 m should be fine.
If you do not use the strobe signal and only show the display at another location, extending the wiring to about 100 m causes almost no problems in practice.
-
Digits that are blank on the display become (F) hex in the BCD output. Is there a solution?
-
On the LDM-110, LDM-210, and LDM-100 series, digits that are blank on the display are output as (F) hex in the BCD output by design.
If you turn off the zero suppression function so that blanks are shown as 0, the BCD data also becomes [0] hex, but the display shows leading zeros.
Alternatively, the receiving side can interpret [F] hex as [0] hex.
On the LDM-110EX and LDM-1000/1000EX, [0] hex is output in the BCD output even when the display is zero-suppressed.
See here for zero suppression ≫
-
What is the BCD output signal when an error is shown on the LDM-110/210?
-
Errors on the LDM-110/210 are shown as "E-" followed by a number, such as "E-12".
In the BCD output signal, "E" is coded as [A] hex, "-" as [B] hex, and blanks as [F] hex.
However, the LDM-110/210 does not output the BCD strobe signal except during correct measurement.
Questions About RS-232C Communication
-
How should the RS-232C cable be wired?
-
The RS-232C pin assignment is shown on the left of the diagram below. Some older models use a 25-pin female D-SUB connector, whose pin assignment is shown on the right.

In the diagram, SG: signal ground, TxD: transmit data, RxD: receive data, RTS: request to send, CTS: clear to send.
In most cases, a commercially available RS-232C cross cable (also called a null modem or reverse cable) can be used, but commercially available cables may not work with some devices.
If you make your own cable, the recommended connections are shown in the table below.Our device Your device SG ---- SG TxD ---- RxD RxD ---- TxD RTS ---- CTS CTS ---- RTS For simpler wiring, leaving RTS-CTS and CTS-RTS unconnected rarely causes problems.
If the recommended connection does not work, in rare cases DSR (data set ready) and DTR (data terminal ready) must be shorted on your device.
-
How far can the RS-232C cable be extended?
-
The standard specifies 15 m or less. If you need a longer cable, convert to an interface suited to long-distance transmission, such as RS-485.
In practice, it works without problems up to about 30 m in most cases, but this is outside the standard, so please note that any problems that occur are not covered by our support.
-
Settings sent to the device via RS-232C sometimes do not work.
-
For commands that use multiple characters, such as setting commands, sending the characters almost simultaneously can cause our device to miss characters and misinterpret the command.
If this happens, the following usually solves the problem.
- Try lowering the baud rate, or send the string one character at a time with a pause between characters.
- For example, when sending "h1234[CR]" ([CR] is the carriage return code) to an LDM-110EX, do not send "h1234[cr]" all at once; instead send "h" – delay – "1" – delay – "2" – delay – "3" – delay – "4" – delay – "[cr]" – delay.
- The required delay depends on the model, but about 50 ms is usually enough.
- If 50 ms does not work, increase the delay. If it still does not work with a 1 s delay, the cause is probably not in our device.
- After a setting command, read back the setting to check that it was set correctly.
- Try lowering the baud rate, or send the string one character at a time with a pause between characters.
-
In RS-232C communication with the LDM-110, I can read the measured value once with [R], but after that communication stops.
-
The following causes are possible.
- You intend to send only "R", but your device automatically adds a carriage return code or similar
- This can happen with older versions of the LDM-110 (version 1.11 or earlier). Newer versions work whether or not a carriage return is added.
- If you use an older version, set your device not to add a carriage return.
- When your device receives data via RS-232C, it automatically echoes characters such as "?" back to our device
- Set your device not to echo characters back to the other device.
- For example, with the ASCII unit of an OMRON PLC, the INPUT# statement echoes "?", so you need to use the LINE INPUT# statement instead.
- You intend to send only "R", but your device automatically adds a carriage return code or similar
-
The display unit shows [E-41], an RS-232C error.
-
This error is almost always caused by mismatched communication settings (baud rate, parity, etc.) between devices.
The most common cause is a baud rate mismatch.
Note that on many devices, changing the baud rate switch does not take effect until the power is turned off and on again.
If you use our automatic start-up unit, the baud rate between the automatic start-up unit and the outer diameter gauge must be 300 bps or 1200 bps.
-
When reading measured values into a PC via RS-232C from the display unit, what sampling rate is possible?
-
It depends on the baud rate setting.
At 9600 bps, sending or receiving one character takes about 1 ms.
For example, reading a measured value from an LDM-110 means reading 8 characters in total: 6 digits, 1 decimal point, and a carriage return code.
In addition, the "R" command must be sent to the LDM-110 to read the measured value, so at 9600 bps the sampling interval is about 10 ms or longer.
At 4800 bps, the sampling interval is twice as long as at 9600 bps.
-
The LDM-110 baud rate setting only goes up to 4800 bps. Can it be set higher?
-
In fact, setting the BAUD switch to 7 sets 9600 bps, and setting it to 8 sets 19200 bps.
The maximum nominal baud rate is 4800 bps because at 9600 bps or higher, delays must be inserted between characters when sending settings via RS-232C.
For reading measured values only, communication works without problems up to 19200 bps.
-
What should I watch out for when using a USB-to-RS-232C converter cable?
-
Many USB-to-RS-232C converter cables are available. Please purchase one that is compatible with the operating system of the PC you use (or plan to use).
We also supply cables, so feel free to contact our Sales Department.
Questions About Outer Diameter Control Units
-
Which is better, extrusion control or take-up control?
-
Comparing the extrusion response and the take-up response to a change in command voltage, the take-up response tends to be better. However, as motor performance has improved, there is now almost no practical difference in control capability.
With extrusion control, the take-up speed is constant, so you can estimate fairly accurately how many hours the line must run from the line speed and production volume, which makes production planning easier.
Extrusion control needs to be used where there is some spare extrusion capacity, whereas take-up control has the advantage that the extruder can run at maximum speed.
Extrusion control is more common in general.
-
How do I choose between the outer diameter control unit models?
-
The DAC+88 with automatic start-up may seem complicated to set up at first, but the start-up function helps improve productivity.
The DAC-96 is the simplest to use, and the DAC-93 is suited to lines with slow line speeds, such as large-diameter cable lines.
The table below shows how to choose between the models.
Product DAC-96 DAC-93 DAC+88 Control method P.I./SPEED control Delay time step control Parameter-setting control Suitable lines General purpose Lines with slow line speeds General purpose (not suited to lines with slow line speeds) Automatic start-up function × ○ Compatible outer diameter gauges Any with an analog deviation output Our products only Controlled equipment Either extrusion or take-up Both extrusion and take-up Speed potentiometer adjustment during AUTO operation Possible Not possible (change the speed by changing the data) Rotary encoder connection Not required Required Not required Data setting for each product Not required Recall a program number External dimensions W144×H72×D191mm
-
Are there any requirements for equipment on which the DAC-96 can be installed?
-
Any type of motor can be used as long as it is equipment whose speed is controlled by a DC voltage command. The polarity of the DC voltage does not matter either.
In rare cases, equipment controls motor speed with an AC slidac; the DAC-96 cannot be installed on such lines.
-
How is the DAC-96 wired?
-
The DAC-96 output +/− is isolated from other circuits and is connected in cascade between the motor signal lines. Note in particular that the − side of the output (pin 6) is not to be connected to signal 0 V.
The basic wiring is described in the DAC-96 instruction manual, so please refer to it.
If you are unsure how to connect it, send us your equipment circuit diagram and we will propose specific wiring based on it.Write "Request for DAC-96 wiring proposal" and the control method (extrusion control or take-up control) on the equipment circuit diagram, and send it to us via our inquiry form or by fax.
Fax: +81-42-645-8938
If you are unsure whether to use extrusion control or take-up control, see this question.
-
What is the output cycle of the DAC-96?
-
The DAC-96 output update cycle is 10 ms (100 times per second).
-
Are there any requirements for equipment on which the DAC+88 can be installed?
-
Any type of motor can be used as long as it is equipment whose speed is controlled by a DC voltage command.
The polarity of the DC voltage does not matter either, but wiring is slightly more complex with negative voltage control.
If the maximum DC command voltage exceeds 10 V, the amplifier option (gain: 2×) is required. If the maximum command voltages of both the extruder and the take-up exceed 10 V, two amplifiers are required.
After the amplifier is installed, the adjustment interval of the potentiometer for raising motor speed changes; setting it to the same position as before doubles the speed.
In rare cases, equipment controls motor speed with an AC slidac; the DAC+88 cannot be installed on such lines.
Questions About Outer Diameter Irregularity Detectors
-
What is the difference between the single-slit and double-slit methods?
-
The diagrams below show a fixed amount of emitted light (shaded area) being blocked by the workpiece (the product being measured).
Single-Slit Method
Detects diameter irregularities by capturing the change in light quantity between when there is no irregularity (left) and when there is one (right).
Movement caused by vibration may be detected as an irregularity, but on the other hand, gently tapered irregularities can also be detected, within limits.
Double-Slit Method
Detects diameter irregularities from the difference in light quantity between the two emitted slits.
There is no difference when there is no irregularity (left); the difference that arises when there is one (right) is used to detect it.
Can also detect irregularities on workpieces that are not perfectly round.
Gradual irregularities may not be detected, but this method is resistant to vibration and suited to fine wire production lines.
-
How is the irregularity level set?
-
The three-digit number on the level adjuster on the unit shows the alarm level in µm or mm. This setting is the absolute difference between the irregular diameter and the normal diameter ((irregular diameter) − (normal diameter)).
If the normal diameter is 1.000 mm and the irregularity level is set to "520", an alarm is output when a lump has a diameter of 1.52 mm or more, or a neck has a diameter of 0.48 mm or less.

Even if the product diameter changes, the irregularity alarm level does not need to be changed as long as the level of irregularity to be detected is the same.
The more sensitive the alarm level, the more demanding the measures against workpiece runout, external vibration, ambient light, water droplets, and so on.
-
What is the relationship between line speed and irregularity width?
-
Line speed and the width of irregularities that can be detected are correlated.
The low cutoff frequency of the outer diameter irregularity detector is about 3 Hz, and the high cutoff frequency is about 8 to 10 kHz (30 kHz or 50 kHz on some models).
Sensitivity therefore tends to drop when the diameter changes very gradually or when an irregularity is very short.
For gradual irregularities, use the following values as a guide to the maximum detectable irregularity width.
Line speed 30 m/min ⇒ up to 15 cm wide
Line speed 60 m/min ⇒ up to 30 cm wide
Line speed 120 m/min ⇒ up to 60 cm wideThe minimum detectable irregularity width depends on the response of the unit and the line speed, and can be calculated with the following formula.

Minimum detectable irregularity width [mm] = length of cable traveled per second [mm/s] ÷ response of the unit [kHz]
However, the minimum irregularity width is limited by the slit width.
-
How well can irregularities in the blind spot of the detection axis be detected?
-
As shown below, when the peak of a lump is tilted relative to the incident light axis, the lump appears narrower, causing a measurement error.

The actual height (h) of a protrusion tilted by α° can be calculated with the following formula.
h=0.5Φ×(cosα-1)+(H×cosα)
This measurement error can be made extremely small by detecting from multiple directions with multiple axes.
Questions About Counting Event Printers
-
I want to print upside down.
-
- Push the locks on both sides of the printer cover inward and remove the cover.
- You can switch the print direction by changing the position of the slide switch on the left side of the board.
-
How does it differ from previous models?
-
See the table below.
Product AD-50 AD-50E AD-500 Number of event messages 10 7 10 Clock pulse rate 2 kpps / 100 pps
(selectable)Up to 2 kpps Up to 2 kpps / 100 pps
(selectable)Reset event message input Front panel switch
Rear terminal block
RS-232CFront panel switch Front panel switch
Rear terminal block
RS-232CChanging message event names Via RS-232C Data setting mode Via RS-232 / USB memory Preset (full / warning) setting Possible Not available Possible Alarm output for settings Yes Yes Initial set digits 6 digits 5 digits 6 digits Lot number digits 4 digits 3 digits 4 digits
(auto-update available)External communication RS-232C Not available RS-232C (standard)
Baud rate: 9600/19200 bps
Questions About Joint Detectors
-
Can a joint detector be used as a break detector?
-
Joint detectors can be used as cable break detectors. However, there are currently several joint detector models, so we select the model after discussing the workpiece and measuring environment with you.
Have a question?
Our team is ready to help.
Contact
Tel. +81-42-645-8920
Hours: 9:00–17:30 JST, Monday–Friday (excluding Japanese public holidays)
Request a Quote
Quotation
Tell us about your application, and we will prepare a quotation.
Request a QuoteBrochures & Manuals
Download
Find the information you need to select and evaluate our products.
Download